Wide Range Reactivity: Mouse Colorimetric Competition ELISA 61.7 ng/mL - 5000 ng/mL Plasma
Catalog No. ABIN6999529
  • Target See all D-Dimer ELISA Kits
    D-Dimer
    Reactivity
    • 9
    • 9
    • 8
    • 4
    • 4
    • 2
    • 2
    • 1
    • 1
    • 1
    • 1
    • 1
    • 1
    Mouse
    Detection Method
    Colorimetric
    Method Type
    Competition ELISA
    Detection Range
    61.7 ng/mL - 5000 ng/mL
    Minimum Detection Limit
    61.7 ng/mL
    Application
    ELISA
    Purpose
    The kit is a wide-range competitive inhibition enzyme immunoassay technique for the in vitro quantitative measurement in various sample types.
    Sample Type
    Plasma
    Analytical Method
    Quantitative
    Specificity
    This assay has high sensitivity and excellent specificity for detection of D-Dimer.
    Sensitivity
    22.2 ng/mL
    Grade
    Wide Range
    Components
    • Pre-coated, ready to use 96-well strip plate, flat buttom
    • Plate sealer for 96 wells
    • Reference Standard
    • Standard Diluent
    • Detection Reagent A
    • Detection Reagent B
    • Assay Diluent A
    • Assay Diluent B
    • Reagent Diluent (if Detection Reagent is lyophilized)
    • TMB Substrate
    • Stop Solution
    • Wash Buffer (30 x concentrate)
    • Instruction manual
    Top Product
    Discover our top product D-Dimer ELISA Kit
  • Sample Volume
    50 µL
    Assay Time
    3 h
    Plate
    Pre-coated
    Protocol
    1. Prepare all reagents, samples and standards,
    2. Add 50μL standard or sample to each well.
      Then add 50μL prepared Detection Reagent A immediately.
      Shake and mix. Incubate 1 hour at 37 °C,
    3. Aspirate and wash 3 times,
    4. Add 100μL prepared Detection Reagent B. Incubate 30 minutes at 37 °C,
    5. Aspirate and wash 5 times,
    6. Add 90μL Substrate Solution. Incubate 10-20 minutes at 37 °C,
    7. Add 50μL Stop Solution. Read at 450 nm immediately.
    Reagent Preparation
    1. Bring all kit components and samples to room temperature (18-25 °C) before use. If the kit will not be used up in one time, please only take out strips and reagents for present experiment, and leave the remaining strips and reagents in required condition.
    2. Standard - Reconstitute the Standard with 1.0mL of Standard Diluent, kept for 10 minutes at room temperature, shake gently (not to foam). The concentration of the standard in the stock solution is 5,000ng/mL. Prepare 5 tubes containing 0.6mL Standard Diluent and produce a triple dilution series. Mix each tube thoroughly before the next transfer. Set up 5 points of diluted standard such as 5,000ng/mL, 1,666.7ng/mL, 555.6ng/mL, 185.2ng/mL, 61.7ng/mL, and the last tubes with Standard Diluent is the blank as 0ng/mL.
    3. Detection Reagent A and Detection Reagent B - Briefly spin or centrifuge the stock Detection A and Detection B before use. Dilute to the working concentration with Assay Diluent A (1:50) and Assay Diluent B (1:100), respectively.
    4. Wash Solution - Dilute 20 mL of Wash Solution concentrate (30x) with 580 mL of deionized or distilled water to prepare 600 mL of Wash Solution (1x).
    5. TMB substrate - Aspirate the needed dosage of the solution with sterilized tips and do not dump the residual solution into the vial again.

    Note:

    1. Making serial dilution in the wells directly is not permitted.
    2. Prepare standard within 15 minutes before assay. Please do not dissolve the reagents at 37 °C directly.
    3. Detection Reagent A and B are sticky solutions, therefore, slowly pipette them to reduce the volume errors.
    4. Please carefully reconstitute Standards or working Detection Reagent A and B according to the instruction, and avoid foaming and mix gently until the crystals are completely dissolved. To minimize imprecision caused by pipetting, use small volumes and ensure that pipettors are calibrated. It is recommended to suck more than 10μL for one pipetting.
    5. The reconstituted Standards, Detection Reagent A and Detection Reagent B can be used only once.
    6. If crystals have formed in the Wash Solution concentrate (30x), warm to room temperature and mix gently until the crystals are completely dissolved.
    7. Contaminated water or container for reagent preparation will influence the detection result.
    Sample Preparation
    • It is recommended to use fresh samples without long storage, otherwise protein degradation and denaturation may occur in these samples, leading to false results. Samples should therefore be stored for a short period at 2 - 8 °C or aliquoted at -20 °C (≤1 month) or -80 °C (≤ 3 months). Repeated freeze-thaw cycles should be avoided. Prior to assay, the frozen samples should be slowly thawed and centrifuged to remove precipitates.
    • If the sample type is not specified in the instructions, a preliminary test is necessary to determine compatibility with the kit.
    • If a lysis buffer is used to prepare tissue homogenates or cell culture supernatant, there is a possibility of causing a deviation due to the introduced chemical substance. The recommended dilution factor is for reference only.
    • Please estimate the concentration of the samples before performing the test. If the values are not in the range of the standard curve, the optimal sample dilution for the particular experiment has to be determined.
    Assay Precision
    Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level of target were tested 20 times on one plate, respectively.
    Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level of target were tested on 3 different plates, 8 replicates in each plate.
    CV(%) = SD/meanX100
    Intra-Assay: CV < 10%
    Inter-Assay: CV < 12%
    Restrictions
    For Research Use only
  • Storage
    4 °C/-20 ° C
    Storage Comment
    1. For unopened kit: All reagents should be stored according to the labels on the vials. The Standard, Detection Reagent A, Detection Reagent B, and 96-well Strip Plate should be stored at -20 °C upon receipt, while the other reagents should be stored at 4 °C.
    2. For opened kits: the remaining reagents must be stored according to the above storage conditions. In addition, please return the unused wells to the foil pouch containing the desiccant and seal the foil pouch with the zipper.
    .
    Expiry Date
    6 months