Phone:
+1 877 302 8632
Fax:
+1 888 205 9894 (Toll-free)
E-Mail:
orders@antibodies-online.com

Cathepsin L ELISA Kit

CTSL1 Reactivity: Mouse Colorimetric Sandwich ELISA 31.25 pg/mL - 2000 pg/mL Cell Culture Supernatant, Plasma, Serum
Catalog No. ABIN6962834
  • Target See all Cathepsin L (CTSL1) ELISA Kits
    Cathepsin L (CTSL1) (Cathepsin L1 (CTSL1))
    Reactivity
    • 7
    • 5
    • 4
    • 2
    • 2
    • 2
    • 2
    • 1
    • 1
    • 1
    • 1
    • 1
    Mouse
    Detection Method
    Colorimetric
    Method Type
    Sandwich ELISA
    Detection Range
    31.25 pg/mL - 2000 pg/mL
    Minimum Detection Limit
    31.25 pg/mL
    Application
    ELISA
    Purpose
    The kit is a sandwich enzyme immunoassay technique for the in vitro quantitative measurement in various sample types.
    Sample Type
    Cell Culture Supernatant, Plasma, Serum
    Analytical Method
    Quantitative
    Specificity
    This kit recognizes Mouse CTSL in samples. No Significant cross-reactivity or interference between Mouse CTSL and analogues was observed.
    Sensitivity
    18.75 pg/mL
    Components
    • Pre-coated, ready to use 96-well strip plate, flat buttom
    • Plate sealer for 96 wells
    • Reference Standard
    • Reference Standard & Sample Diluent
    • Biotinylated Detection Antibody (100 x concentrate)
    • HRP Conjugate (100 x concentrate)
    • Biotinylated Detection Antibody Diluent
    • HRP Conjugate Diluent
    • Substrate Reagent
    • Stop Solution
    • Wash Buffer (25 x concentrate)
    • Instruction manual
    Featured
    Discover our best selling CTSL1 ELISA Kit
    Top Product
    Discover our top product CTSL1 ELISA Kit
  • Sample Volume
    100 μL
    Assay Time
    3.5 h
    Plate
    Pre-coated
    Protocol
    1. Add 100 µL standard or sample to each well. Incubate for 90 min at 37 °C.
    2. Remove the liquid. Add 100 µL Biotinylated Detection Antibody. Incubate for 1 hour at 37 °C.
    3. Aspirate and wash 3 times.
    4. Add 100 µL HRP Conjugate. Incubate for 30 min at 37 °C.
    5. Aspirate and wash 5 times.
    6. Add 90 µL Substrate Reagent. Incubate for 15 min at 37 °C.
    7. Add 50 µL Stop Solution. Read at 450 nm immediately.
    8. Calculation of results.
    Reagent Preparation
    1. Bring all reagents to room temperature (18~25 °C) before use. Follow the Microplate reader manual for set-up and preheat it for 15 min before OD measurement.
    2. Wash Buffer: Dilute 30 mL of Concentrated Wash Buffer with 720 mL of deionized or distilled water to prepare 750 mL of Wash Buffer.Note: if crystals have formed in the concentrate, warm it in a 40 °C water bath and mix it gently until the crystals have completely dissolved
    3. Standard working solution: Centrifuge the standard at 10,000xg for 1 min. Add 1.0 mL of Reference Standard &Sample Diluent, let it stand for 10 min and invert it gently several times. After it dissolves fully, mix it thoroughly with a pipette. This reconstitution produces a working solution of 2000 pg/mL. Then make serial dilutions as needed. The recommended dilution gradient is as follows: 2000, 1000, 500, 250, 125, 62.5, 31.25, 0 pg/mL. Dilution method: Take 7 EP tubes, add 500 μLof Reference Standard & Sample Diluent to each tube. Pipette 500 μLof the 2000 pg/mL working solution to the first tube and mix up to produce a 1000 pg/mL working solution. Pipette 500 μLof the solution from the former tube into the latter one according to these steps. The illustration below is for reference. Note: the last tube is regarded as a blank. Don't pipette solution into it from the former tube.
    4. Biotinylated Detection Antibody working solution: Calculate the required amount before the experiment (100 μL/well). In preparation, slightly more than calculated should be prepared. Centrifuge the stock tube before use, dilute the 100x Concentrated Biotinylated Detection Antibody to 1xworking solution with Biotinylated Detection Antibody Diluent.
    5. Concentrated HRP Conjugate working solution: Calculate the required amount before the experiment (100 μL/well). In preparation, slightly more than calculated should be prepared. Dilute the 100x Concentrated HRP Conjugate to 1x working solution with Concentrated HRP Conjugate Diluent.
    Sample Preparation
    • It is recommended to use fresh samples without long storage, otherwise protein degradation and denaturation may occur in these samples, leading to false results. Samples should therefore be stored for a short period at 2 - 8 °C or aliquoted at -20 °C (≤1 month) or -80 °C (≤ 3 months). Repeated freeze-thaw cycles should be avoided. Prior to assay, the frozen samples should be slowly thawed and centrifuged to remove precipitates.
    • If the sample type is not specified in the instructions, a preliminary test is necessary to determine compatibility with the kit.
    • If a lysis buffer is used to prepare tissue homogenates or cell culture supernatant, there is a possibility of causing a deviation due to the introduced chemical substance. The recommended dilution factor is for reference only.
    • Please estimate the concentration of the samples before performing the test. If the values are not in the range of the standard curve, the optimal sample dilution for the particular experiment has to be determined.
    Assay Precision
    Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level Mouse CTSL were tested 20 times on one plate, respectively.
    Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level Mouse CTSL were tested on 3 different plates, 20 replicates in each plate.
    Both intra-CV and inter-CV are < 10 %.
    Restrictions
    For Research Use only
  • Storage
    4 °C,-20 °C
    Storage Comment
    1. For unopened kit: All reagents should be stored according to the labels on the vials, so they are stable up to 12 months after receipt of the kit. The Reference Standard, Biotinylated Detection Antibody, HRP Conjugate and the 96-well stripe plate should be stored at -20 °C upon receipt while the other reagents should be stored at 4 °C.
    2. For used kit: When the kit is used, the remaining reagents need to be stored according to the above storage condition. Besides, please return the unused wells to the foil pouch containing the desiccant pack, and zip-seal the foil pouch.
    Expiry Date
    12 months
  • Target See all Cathepsin L (CTSL1) ELISA Kits
    Cathepsin L (CTSL1) (Cathepsin L1 (CTSL1))
    Alternative Name
    Cathepsin L (CTSL1 Products)
    Synonyms
    CATL ELISA Kit, CTSL1 ELISA Kit, MEP ELISA Kit, 1190035F06Rik ELISA Kit, Ctsl1 ELISA Kit, fs ELISA Kit, nkt ELISA Kit, CATHL ELISA Kit, CatL ELISA Kit, Ctsl ELISA Kit, CTSL ELISA Kit, mep ELISA Kit, catl ELISA Kit, ctsl ELISA Kit, cpl-1 ELISA Kit, MGC80629 ELISA Kit, MGC107932 ELISA Kit, LOC663234 ELISA Kit, 26-29-p ELISA Kit, Cp1 ELISA Kit, si:dkey-121a11.6 ELISA Kit, CTSV ELISA Kit, catL ELISA Kit, cb15 ELISA Kit, ctsl1b ELISA Kit, hgg1 ELISA Kit, wu:fb30g09 ELISA Kit, zgc:92089 ELISA Kit, cathepsin L ELISA Kit, cathepsin V ELISA Kit, cathepsin L L homeolog ELISA Kit, cathepsin L1 ELISA Kit, Cathepsin L1 ELISA Kit, Cathepsin L ELISA Kit, testin gene ELISA Kit, cathepsin 12 ELISA Kit, cathepsin Lb ELISA Kit, cathepsin L.1 ELISA Kit, CTSL ELISA Kit, Ctsl ELISA Kit, CTSV ELISA Kit, ctsl.L ELISA Kit, ctsl ELISA Kit, LOC100008830 ELISA Kit, catl1 ELISA Kit, 26-29-p ELISA Kit, catl ELISA Kit, Tsp_11596 ELISA Kit, Testin ELISA Kit, cts12 ELISA Kit, LOC102156614 ELISA Kit, ctslb ELISA Kit, ctsl.1 ELISA Kit
    Background
    CTSL1, CATL, MEP
    Pathways
    Activation of Innate immune Response, Toll-Like Receptors Cascades
You are here: