INHBB ELISA Kit
-
- Target See all INHBB ELISA Kits
- INHBB (Inhibin, beta B (INHBB))
-
Reactivity
- Mouse
- Detection Method
- Colorimetric
- Method Type
- Sandwich ELISA
- Detection Range
- 15.625 pg/mL - 1000 pg/mL
- Minimum Detection Limit
- 15.625 pg/mL
- Application
- ELISA
- Purpose
- The kit is a sandwich enzyme immunoassay for the in vitro quantitative measurement of INHbB in mouse serum, plasma and other biological fluids.
- Sample Type
- Plasma, Serum
- Analytical Method
- Quantitative
- Specificity
- This assay has high sensitivity and excellent specificity for detection of this index.
- Cross-Reactivity (Details)
- No significant cross-reactivity or interference between this index and analogues was observed. Note: Limited by current skills and knowledge, it is impossible for us to complete the cross- reactivity detection between this index and all the analogues, therefore, cross reaction may still exist.
- Sensitivity
- 5.9 pg/mL
- Components
-
- Pre-coated, ready to use 96-well strip plate
- Standard (freeze dried)
- Standard Diluent
- Detection Reagent A
- Detection Reagent B
- Assay Diluent A
- Assay Diluent B
- TMB
- Stop Solution
- Wash Buffer (30X)
- Plate sealer for 96 wells
- Instruction manual
- Material not included
-
- Microplate reader with 450 ± 10nm filter.
- Precision single or multi-channel pipettes and disposable tips.
- Eppendorf Tubes for diluting samples.
- Deionized or distilled water.
- Absorbent paper for blotting the microtiter plate.
- Container for Wash Solution.
- Featured
- Discover our best selling INHBB ELISA Kit
- Top Product
- Discover our top product INHBB ELISA Kit
-
-
- Sample Volume
- 100 μL
- Assay Time
- 1 - 4.5 h
- Plate
- Pre-coated
- Protocol
-
1. Prepare all reagents, samples and standards
2. Add 100µL standard or sample to each well. Incubate 2 hours at 37°C
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C
4. Aspirate and wash 3 times
5. Add 100µL prepared Detection Reagent B. Incubate 1 hour at 37°C
6. Aspirate and wash 5 times
7. Add 90µL Substrate Solution. Incubate 15-25 minutes at 37°C
8. Add 50µL Stop Solution. Read at 450nm immediately. - Assay Procedure
-
The microtiter plate provided in this kit has been pre-coated with an antibody specific to the index. Standards or samples are then added to the appropriate microtiter plate wells with a biotin-conjugated antibody preparation specific to the index. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain the index, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of the index in the samples is then determined by comparing the O.D. of the samples to the standard curve.
- Assay Precision
-
- Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level the index were tested 20 times on one plate, respectively.
- Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level the index were tested on 3 different plates, 8 replicates in each plate.
- CV(%) = SD/meanX100
- Intra-assay: CV<10%
- Inter-assay: CV<12%
- Restrictions
- For Research Use only
-
- Precaution of Use
- The Stop Solution suggested for use with this kit is an acid solution. Wear eye, hand, face, and clothing protection when using this material.
- Handling Advice
- The stability of ELISA kit is determined by the loss rate of activity. The loss rate of this kit is less than 5 % within the expiration date under appropriate storage conditions. Note: To minimize unnecessary influences on the performance, operation procedures and lab conditions, especially room temperature, air humidity and incubator temperatures should be strictly regulated. It is also strongly suggested that the whole assay is performed by the same experimenter from the beginning to the end.
- Storage
- 4 °C,-20 °C
- Storage Comment
- The Assay Plate, Standard, Detection Reagent A and Detection Reagent B should be stored at -20°C upon being received. After receiving the kit , Substrate should be always stored at 4°C.Other reagents are kept according to the labels on vials. But for long term storage, please keep the whole kit at -20°C. The unused strips should be kept in a sealed bag with the desiccant provided to minimize exposure to damp air. The test kit may be used throughout the expiration date of the kit (six months from the date of manufacture). Opened test kits will remain stable until the expiring date shown, provided it is stored as prescribed above.
- Expiry Date
- 12 months
-
- Target See all INHBB ELISA Kits
- INHBB (Inhibin, beta B (INHBB))
- Alternative Name
- INHbB (INHBB Products)
- Synonyms
- actbb ELISA Kit, actbetaB ELISA Kit, inhibin ELISA Kit, inhibin, beta B ELISA Kit, inhibin beta B L homeolog ELISA Kit, inhibin beta B subunit ELISA Kit, inhibin beta B ELISA Kit, inhibin beta-B ELISA Kit, inhibin beta B chain ELISA Kit, inhbb ELISA Kit, inhbb.L ELISA Kit, INHBB ELISA Kit, Inhbb ELISA Kit, LOC100719829 ELISA Kit
- Background
- Alternative name: INH-BB, Activin AB Beta Polypeptide
- Gene ID
- 16324
- UniProt
- Q04999
- Pathways
- Peptide Hormone Metabolism, Hormone Activity, Negative Regulation of Hormone Secretion
-